Product Recalls in Virginia

Product recalls affecting Virginia — including food, drugs, consumer products, medical devices, and vehicles distributed to Virginia. RecallDetector monitors distribution data from the FDA, USDA, CPSC, and NHTSA to show you which recalls are relevant to your area. 1,773 recalls have been distributed to Virginia in the last 12 months.

53,829 total recalls
1,773 in last 12 months

Showing 4416144180 of 53,829 recalls

Medical DeviceNovember 4, 2014· Aircraft Medical Limited

Recalled Item: McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items...

The Issue: The blade retaining clip may be difficult to engage and thereby increase the...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceNovember 4, 2014· Novadaq Technologies, Inc.

Recalled Item: Non-sterile drape Recalled by Novadaq Technologies, Inc. Due to Mislabeling

The Issue: Non sterile-drape LUNA drape was inadvertently mislabeled with an additional...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class III — Low RiskOngoing
Medical DeviceNovember 4, 2014· Novadaq Technologies, Inc.

Recalled Item: LUNA Chest Recalled by Novadaq Technologies, Inc. Due to Mislabeling

The Issue: Non sterile-drape LUNA drape was inadvertently mislabeled with an additional...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class III — Low RiskOngoing
Medical DeviceNovember 4, 2014· Medtronic MiniMed Inc.

Recalled Item: Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets Recalled by...

The Issue: Tubing may detach at the connect/disconnect location on the MiniMed...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceNovember 4, 2014· Smith & Nephew, Inc.

Recalled Item: LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT Recalled by Smith & Nephew,...

The Issue: The stem taper on this device is undersized; the gauge point is not deep enough.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceNovember 4, 2014· Smith & Nephew, Inc.

Recalled Item: LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT Recalled by Smith & Nephew,...

The Issue: The stem taper on this device is undersized; the gauge point is not deep enough.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceNovember 3, 2014· Elekta, Inc.

Recalled Item: MOSAIQ Product Usage: MOSAIQ is an oncology information system used Recalled...

The Issue: A problem can exist in MOSAIQ resulting in the display of incorrect numeric...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
DrugNovember 3, 2014· REFA Enterprises, LLC

Recalled Item: Forever Beautiful INFINITY Recalled by Enterprises, LLC Due to Marketed...

The Issue: Marketed Without an Approved NDA/ANDA: FDA analysis found product to contain...

Recommended Action: Stop taking this medication. Contact your pharmacy or healthcare provider.

Class I — Serious Health HazardOngoing
Refund
DrugNovember 3, 2014· REFA Enterprises, LLC

Recalled Item: Forever Beautiful BEE POLLEN Recalled by Enterprises, LLC Due to Marketed...

The Issue: Marketed Without an Approved NDA/ANDA: FDA analysis found product to contain...

Recommended Action: Stop taking this medication. Contact your pharmacy or healthcare provider.

Class I — Serious Health HazardOngoing
Refund
FoodOctober 31, 2014· Nature's Way Purewater Systems, Inc.

Recalled Item: ESSENTIA Hydration Perfected Purified Water and Electrolytes Recalled by...

The Issue: Product may contain mold.

Recommended Action: Do not consume. Return to store for a refund or discard.

Class II — Potential Health HazardOngoing
Refund
DrugOctober 30, 2014· Baxter Healthcare Corp.

Recalled Item: Heparin Sodium in 0.9% Sodium Chloride Injection Recalled by Baxter...

The Issue: Subpotent Drug: Heparin raw material was found to have low potency

Recommended Action: Stop taking this medication. Contact your pharmacy or healthcare provider.

Class II — Potential Health HazardOngoing
Refund
Medical DeviceOctober 30, 2014· Alcon Research, Ltd.

Recalled Item: INFINITI VISION SYSTEM ULTRASOUND FMS. Single-use fluid management system....

The Issue: Alcon is conducting a medical device recall due to the possible presence of...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceOctober 29, 2014· Remel Inc

Recalled Item: Remel TB Potassium Permanganate (.5% Aqueous) 40192 Recalled by Remel Inc...

The Issue: The product may appear cloudy or discolored and may not perform correctly in...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceOctober 29, 2014· Radiometer America Inc

Recalled Item: Crea A and B membranes Recalled by Radiometer America Inc Due to Negative...

The Issue: Negative drift on QC during in-use lifetime of CREA membranes can be observed.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceOctober 29, 2014· Respironics California Inc

Recalled Item: Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit...

The Issue: If the power supply fan mounting screws are installed incorrectly, the ends...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
FoodOctober 29, 2014· Mt. Olive Pickle Company Inc

Recalled Item: Mt. Olive Recalled by Mt. Olive Pickle Company Inc Due to Product may...

The Issue: Product may contain glass fragments.

Recommended Action: Do not consume. Return to store for a refund or discard.

Class II — Potential Health HazardOngoing
Refund
FoodOctober 28, 2014· Marathon Ventures Inc

Recalled Item: Macadamia nuts in various formats: (1.) Western Family Macadamia Diced Nuts...

The Issue: Macadamia nuts may have possible contamination with Salmonella florida.

Recommended Action: Do not consume. Return to store for a refund or discard.

Class I — Serious Health HazardOngoing
Refund
Medical DeviceOctober 28, 2014· Cooper Vision Caribbean Corp.

Recalled Item: Soft Contact Lens Recalled by Cooper Vision Caribbean Corp. Due to Portions...

The Issue: Portions of the lots may contain units with an incorrect lens axis condition.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceOctober 28, 2014· Summit Industries Inc.

Recalled Item: The J700 Floor Mounted Tube Stand (FMTS) is intended to Recalled by Summit...

The Issue: The welds on J700 tube stands may be insufficient resulting in tube stand...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceOctober 28, 2014· Siemens Medical Solutions USA, Inc

Recalled Item: Mammomat Inspiration system: Product Usage: mammography exams Recalled by...

The Issue: It was determined that if the Mammomat Inspiration system is not secured to...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing