Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 Recalled by Respironics California Inc Due to If the power supply fan mounting screws are...

Date: October 29, 2014
Company: Respironics California Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Respironics California Inc directly.

Affected Products

Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators are microprocessor-controlled, electrically powered mechanical ventilators. Intended for use by medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients.

Quantity: 3,943 units total (3,071 units in US)

Why Was This Recalled?

If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Respironics California Inc

Respironics California Inc has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report