Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Mammomat Inspiration system: Product Usage: mammography exams Recalled by Siemens Medical Solutions USA, Inc Due to It was determined that if the Mammomat Inspiration...

Date: October 28, 2014
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Quantity: 3

Why Was This Recalled?

It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.

Where Was This Sold?

This product was distributed nationwide across all 50 states.