Product recalls affecting Rhode Island — including food, drugs, consumer products, medical devices, and vehicles distributed to Rhode Island. RecallDetector monitors distribution data from the FDA, USDA, CPSC, and NHTSA to show you which recalls are relevant to your area. 2,496 recalls have been distributed to Rhode Island in the last 12 months.

47,970 total recalls
2,496 in last 12 months

Showing 1092110940 of 27,456 recalls

Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: 76 cm (30") Appx 3.3 ml Recalled by ICU Medical, Inc. Due to Potential for...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· Vascular Solutions, Inc.

Recalled Item: Langston Dual Lumen Catheter. Recalled by Vascular Solutions, Inc. Due to...

The Issue: The inner lumen of the Langston catheter may separate from the hub during...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceAugust 31, 2020· Brainlab AG

Recalled Item: Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for...

The Issue: Brainlab Ultrasound Navigation Software does not support the modification of...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: Spiros CLOSED MALE LUER w/RED CAP Recalled by ICU Medical, Inc. Due to...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· Vascular Solutions, Inc.

Recalled Item: Langston Dual Lumen Catheter. Recalled by Vascular Solutions, Inc. Due to...

The Issue: The inner lumen of the Langston catheter may separate from the hub during...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: 42 IN (107 cm) Appx 5.0 ml Recalled by ICU Medical, Inc. Due to Potential...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: Appx 1.7 ml Bag Spike w/Integrated Clave Recalled by ICU Medical, Inc. Due...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: 5 IN (13 cm) Appx 1.5 ml Recalled by ICU Medical, Inc. Due to Potential for...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· Olympus Corporation of the Americas

Recalled Item: Olympus Medical Systems Corporation (OMSC) Evis Exera II Bronchovideoscope...

The Issue: The BF-Q180 bronchoscope is being recalled because it does not have a 510(k)...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: 30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 31, 2020· ICU Medical, Inc.

Recalled Item: 30" (76 cm) Appx 3.2 ml Recalled by ICU Medical, Inc. Due to Potential for...

The Issue: Potential for spinning versions of male luers to exhibit leaks due to a...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 28, 2020· Siemens Healthcare Diagnostics Inc

Recalled Item: Siemens epoc BGEM Test Card-In vitro diagnostic device for the Recalled by...

The Issue: Sporadically inconsistent discrepant (low bias) glucose results on card lot...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 28, 2020· Greiner Bio-One North America, Inc.

Recalled Item: VACUETTE TUBE 5 mL CAT Serum Separator Clot Activator Recalled by Greiner...

The Issue: Blood collection tubes experienced an issue with separation and clotting

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 27, 2020· Roche Diagnostics Operations, Inc.

Recalled Item: Tina-quant Complement C4 ver.2 Recalled by Roche Diagnostics Operations,...

The Issue: Roche has confirmed customer complaints of low quality (QC) recovery and...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class III — Low RiskOngoing
Medical DeviceAugust 27, 2020· Roche Diagnostics Operations, Inc.

Recalled Item: Albumin Gen.2 Recalled by Roche Diagnostics Operations, Inc. Due to Roche...

The Issue: Roche has confirmed customer complaints of low quality (QC) recovery and...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class III — Low RiskOngoing
Medical DeviceAugust 26, 2020· Beckman Coulter Inc.

Recalled Item: BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol...

The Issue: Due to alkaline phosphatase (ALP) associated interference causing falsely...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 26, 2020· ev3 Inc.

Recalled Item: Pacific Xtreme (PTCA Balloon Dilation Catheter) Recalled by ev3 Inc. Due to...

The Issue: Labeling discrepancy for the Rated Burst Pressure (RBP) value.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 25, 2020· Terumo Cardiovascular Systems Corporation

Recalled Item: Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32 Recalled...

The Issue: Internal testing at the supplier, revealed that a crimped wire was placed...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 25, 2020· Genicon, Inc.

Recalled Item: Aquas PowerFlow Tube Set - Product Usage: The device delivers Recalled by...

The Issue: The product design incorporating a motor driven impeller pump powered by one...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceAugust 25, 2020· Medtronic Vascular

Recalled Item: Rashkind Balloon Septostomy Catheter Recalled by Medtronic Vascular Due to...

The Issue: Subsequent failure of catheters in the field and failed quality testing

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing