Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

42 IN (107 cm) Appx 5.0 ml Recalled by ICU Medical, Inc. Due to Potential for spinning versions of male luers to...

Date: August 31, 2020
Company: ICU Medical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ICU Medical, Inc. directly.

Affected Products

42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4.4 ml, Ext Set w/MicroClave, 0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3381; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Drop-in Red Cap, 2 Hanger, REF: CH3386; 40" (102 cm) Appx 7.3 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: CH3401; 60" (152 cm) Appx 0.62 ml, Smallbore Ext Set w/MicroClave, Clamp, Spiros w/Red Cap, REF: CH3405; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, REF: CH3453; 73" (185 cm) Appx 2.1 ml, Smallbore Ext Set w/MicroClave Clear, Spiros, Clamp, REF: CH3485; 42" (107 cm) Appx 4.8 ml, 15 Drop Admin Set w/Spiros w/Red Cap, Bag Hanger, Drop-In Priming Cap, REF: CH3517; 8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/MicroClave Clear, 2 Spiros, 3 Clamps, Priming Cap, REF: CH3523; 74" (188 cm) Smallbore Ext Set w/MicroClave, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3533; 74 IN (188 cm) Appx 0.92 ml, Smallbore Set, MicroClave, 1.2 Micron Fil, Spiros, Red Cap, REF: CH3538; 18" (46 cm) Appx 5.1 ml, Ext Set w/0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3556; 60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; 4-Way Clave Stopcock w/2 Spiros, REF: B4194; Spiros, Closed Male Connector w/Red Cap (10 Units), REF: Z7093; 47" (119 cm) 150 ml Burette Set w/MicroClave, 20 Drop In-Line Drip Chamber w/Ball Shut Off, Spiros, REF: Z7182

Quantity: 19850

Why Was This Recalled?

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ICU Medical, Inc.

ICU Medical, Inc. has 592 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report