Product Recalls in Vermont

Product recalls affecting Vermont — including food, drugs, consumer products, medical devices, and vehicles distributed to Vermont. RecallDetector monitors distribution data from the FDA, USDA, CPSC, and NHTSA to show you which recalls are relevant to your area. 2,515 recalls have been distributed to Vermont in the last 12 months.

47,027 total recalls
2,515 in last 12 months

Showing 58815900 of 27,157 recalls

Medical DeviceJune 5, 2023· ORTHO-CLINICAL DIAGNOSTICS

Recalled Item: VITROS Immunodiagnostic Products Troponin I ES Reagent Pack Recalled by...

The Issue: Reagent packs could potentially contain incorrect wells, which will generate...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· ORTHO-CLINICAL DIAGNOSTICS

Recalled Item: VITROS Immunodiagnostic Products Troponin I ES Calibrators Recalled by...

The Issue: Reagent packs could potentially contain incorrect wells, which will generate...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· Datascope Corp.

Recalled Item: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) Recalled by Datascope...

The Issue: IABP may lose the ability to charge batteries in one or both bay slots....

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 5, 2023· Siemens Medical Solutions USA, Inc

Recalled Item: Cios Select VA 21 with FD- A Mobile X-ray system Recalled by Siemens Medical...

The Issue: Incorrect glue was applied on the plexiglass sheet on the single tank cover...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· Datascope Corp.

Recalled Item: Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) Recalled by Datascope...

The Issue: IABP may lose the ability to charge batteries in one or both bay slots....

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 5, 2023· Datascope Corp.

Recalled Item: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) Recalled by Datascope...

The Issue: An unexpected shutdown of the IABP may occur due to failures of tantalum...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 5, 2023· Datascope Corp.

Recalled Item: Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) Recalled by Datascope...

The Issue: An unexpected shutdown of the IABP may occur due to failures of tantalum...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 5, 2023· B. Braun Medical, Inc.

Recalled Item: Design Options¿ / Epidural anesthesia kit (10 count carton) Recalled by B....

The Issue: Kits were assembled with the incorrect Filter Straw.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· Luminex Corporation

Recalled Item: VERIGENE Clostridium difficile Nucleic Acid Test Recalled by Luminex...

The Issue: It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· Luminex Corporation

Recalled Item: Verigene Enteric Pathogens Nucleic Acid Test (EP) Recalled by Luminex...

The Issue: It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 5, 2023· B. Braun Medical, Inc.

Recalled Item: Perifix¿ / Epidural anesthesia kit (10 count carton) Recalled by B. Braun...

The Issue: Kits were assembled with the incorrect Filter Straw.

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 2, 2023· Covidien, LLC

Recalled Item: Covidien Cytosponge Cell Collection Device Recalled by Covidien, LLC Due to...

The Issue: Cell Collection Device may be at increased risk of the sponge detaching from...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class II — Potential Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Universal Dual Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Universal Plus Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA 2000 Patient Return Electrode. used during electrosurgery...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Reusable Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Universal Plus Dual Patient Return Electrode. used...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Dual Reusable Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing
Medical DeviceJune 1, 2023· Megadyne Medical Products, Inc.

Recalled Item: MEGADYNE MEGA SOFT Universal Patient Return Electrode. used during...

The Issue: Firm has received reports of patient burns in surgical procedures where...

Recommended Action: Stop using. Contact the manufacturer or your healthcare provider.

Class I — Serious Health HazardOngoing