Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

3:1 Dermacarrier Recalled by Zimmer Surgical Inc Due to Due to manufacturing issue, the product is labeled...

Date: January 2, 2024
Company: Zimmer Surgical Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer Surgical Inc directly.

Affected Products

3:1 Dermacarrier, Model Number 00219501300, skin graft carrier

Quantity: 76 boxes (1,520 units) US; 521 boxes (10,420 units) OUS

Why Was This Recalled?

Due to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer Surgical Inc

Zimmer Surgical Inc has 26 total recalls tracked by RecallDetector.

Related Recalls