Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Zimmer Trabecular Metal Total Ankle Talar Component Right Sizes 1-6 Recalled by Zimmer, Inc. Due to As part of routine manufacturing process monitoring by...

Date: November 12, 2013
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Zimmer Trabecular Metal Total Ankle Talar Component Right Sizes 1-6 Total ankle arthroplasty is intended to provide a patient with limited mobility by restoring alignment, reducing pain and preserving the flexion/extension movement within the ankle joint.The Zimmer Trabecular Metal Total Ankle is indicated as a total ankle replacement in primary or revision surgery for patients with: " Rheumatoid arthritis. " Post-traumatic arthritis. " Degenerative arthritis. This device is intended for cemented use only.

Quantity: 265 units (US) 114 units (outside US)

Why Was This Recalled?

As part of routine manufacturing process monitoring by Zimmer, it was discovered that the Tibial Base and Talar Components from the Total Ankle system have the potential of containing elevated levels of manufacturing materials residue on the implants. The specific material was found to be mass media utilized during the manufacturing process to provide the uniform surface finish. This mass media

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls