Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Zimmer Title 2 Bone Screw Remover Instrument Recalled by Zimmer, Inc. Due to Possibility of the weld to the quick-connect breaking...

Date: August 26, 2013
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial Spinal System, used for posterior lumbar fusion procedures.

Quantity: 14 devices

Why Was This Recalled?

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report