Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem Recalled by Zimmer, Inc. Due to PMMA coating does not meet specifications. Area of...

Date: February 5, 2015
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem, Smooth Stem Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented.

Quantity: 1 unit

Why Was This Recalled?

PMMA coating does not meet specifications. Area of the stem may be missing PMMA coating.

Where Was This Sold?

This product was distributed to 1 state: TX

Affected (1 state)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report