Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Trilogy F/M Acetabular Shell without holes SOLID HA. Several ODs: 38MM Recalled by Zimmer, Inc. Due to Some units of Trilogy shells manufactured since March...

Date: September 14, 2012
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Trilogy F/M Acetabular Shell without holes SOLID HA. Several ODs: 38MM, 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

Quantity: 275,952 units

Why Was This Recalled?

Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected