Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Persona The Personalized Knee System Recalled by Zimmer, Inc. Due to All lots of the Persona Cemented Tibial Drill,...

Date: November 14, 2012
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE G, LEFT TIBIA SIZE G RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.

Quantity: 346

Why Was This Recalled?

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.