Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Persona The Personalized Knee System 2.5 mm Female Screw 48 Recalled by Zimmer, Inc. Due to The Persona 48mm X 2.5mm Female Screw is...

Date: September 9, 2014
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Persona The Personalized Knee System 2.5 mm Female Screw 48 mm Length Sterile Qty-2. Sterilized using irradiation, single use.

Quantity: 16,380 units

Why Was This Recalled?

The Persona 48mm X 2.5mm Female Screw is being recalled due to potential use of the screw in cortical bone, and corresponding stripping of the 2.5mm Persona Hex Driver.

Where Was This Sold?

This product was distributed nationwide across all 50 states.