Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Osteobond Bone Cement Bone cement monomer is kitted with bone Recalled by Zimmer, Inc. Due to Samples of product from Supplier lot 14D0809 and...

Date: December 8, 2014
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants.

Quantity: 40 units

Why Was This Recalled?

Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet the requirements of ZWI 43.109 and corresponding visual aid 55-0000-310-01.

Where Was This Sold?

This product was distributed to 3 states: CA, KS, MI

Affected (3 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls