Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Moore Hip Prosthesis Long Fenestrated Stem 38 mm (1-1/2 Inch) Recalled by Zimmer, Inc. Due to Zimmer is initiating a voluntary recall of 20...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.
Affected Products
Moore Hip Prosthesis Long Fenestrated Stem 38 mm (1-1/2 Inch) Head Diameter 203 mm Length. Product Usage: The Moore Hip Prosthesis is a femoral him stem that can be used where there is roentgenogrpahic evidence of a satisfactory actabulum and sufficient bone in the femoral neck to seat the prosthesis with or without bone cement.
Quantity: 68
Why Was This Recalled?
Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Zimmer, Inc.
Zimmer, Inc. has 437 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report