Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

M/DN R-F IM NAIL 9MM DIA Recalled by Zimmer, Inc. Due to Zimmer recently conducted a review of all historic...

Date: October 11, 2013
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

M/DN R-F IM NAIL 9MM DIA , 10 MM, 11 MM, 12 MM, 13 MM and 14 MM. Item codes starting with 0022400

Quantity: 7,044,680 total

Why Was This Recalled?

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Where Was This Sold?

This product was distributed nationwide across all 50 states.