Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Knee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM Recalled by Zimmer, Inc. Due to Zimmer is recalling sterile-packaged implants and instruments packaged...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.
Affected Products
Knee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANT 24MMDX145MM SHARP FLUTED ST 20MMDX75MM SHARP FLUTED ST 22MMDX75MM SHARP FLTD ST 24MMDX75MM SHARP FLUTED ST 20MMDX130MM SHARP FLUTED ST 22MMDX130MM SHARP FLUTED ST 24MMDX130MM CEMENTED STEM EXT 13MM X 145MM IMPACTOR SLEEVE* NGK STEM EXT 12X245MM (200MM) NGK STEM EXT 13X245MM (200MM) NGK STEM EXT 14X245MM (200MM) NGK STEM EXT 15X245MM (200MM) NGK STEM EXT 16X245MM (200MM) NGK STEM EXT 17X245MM (200MM) NGK STEM EXT 18X245MM (200MM) NGK STEM EXT 20X245MM (200MM) NGK STEM EXT 22X245MM (200MM) SMOOTH REV STM 9.5MM X 19 SMOOTH REV STM 11.5MM X 1 SMOOTH REV STM 13.5MM X 1 SMOOTH REV STM 15.5MM X 1 SMOOTH REV STM 17.5MM X 1 REV FEM STM 10.5MM X 200M REV FEM STM 12.5MM X 200M REV FEM STM 14.5MM X 200M REV FEM STM 16.5MM X 200M 18.5X200MM REV STEM REV FEM STM 10.5MM X 250M REV FEM STM 12.5MM X 250M REV FEM STM 14.5MM X 250M REV FEM STM 16.5MM X 250M 12.5X155MM FLUTED OFFST R 14.5X155MM FLUTED OFFST R 16.5X155MM FLUTED OFFST R 18.5X155MM FLUTED OFFST R 20.5X155MM FLUTED OFFST R 12.5X175MM FLUTED REV STM FLUTE REV STEM - 14.5MM X FLUTE REV STEM - 16.5MM X FLUTE REV STEM - 18.5MM X FLUTE REV STEM - 20.5MM X FLUTE REV STEM - 22.5MM X FLUTED REV STM 10.5MM X 1 FLUTED REV STM 11.5MM X 1 FLUTED REV STM 12.5MM X 1 FLUTED REV STM 13.5MM X 1 FLUTED REV STM 14.5MM X 1 FLUTED REV STM 16.5MM X 1 FLUTED REV STM 18.5MM X 1
Quantity: 972,125 units
Why Was This Recalled?
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
Where Was This Sold?
Worldwide Distribution.
About Zimmer, Inc.
Zimmer, Inc. has 437 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report