Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Bigliani/Flatow¿ The Complete Shoulder Solution Recalled by Zimmer, Inc. Due to The firm is initiating a removal of one...

Date: June 11, 2013
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Bigliani/Flatow¿ The Complete Shoulder Solution, Humeral Stem Provisional 12mm x 170mm This reusable trial device is used in orthopaedic joint replacement and is specific to the replaced joint (e.g. hip, knee, shoulder, ankle, elbow). It can be made of metal or plastic.

Quantity: 1

Why Was This Recalled?

The firm is initiating a removal of one lot of the Bigliani/Flatow Humeral Provisional Stem (00-4301-012-17, lot 62283991) as the stems manufactured under the lot are 14mm x 170mm devices incorrectly etched and packaged as 12mm x 170mm devices.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls