Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB Recalled by Zimmer, Inc. Due to A complaint was received for a discolored All...

Date: February 25, 2014
Company: Zimmer, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.

Affected Products

Apollo" Knee System: PS TIB INS SZ0/16MM APOLL PS TIB INS SZ0/19MM APOLL PS TIB INS SZ0/22MM APOLL PS TIB INS SZ1/16MM APOLL PS TIB INS SZ2/16M APOLLO PS TIB INS SZ2/22MM APOLL PS TIB INS SZ3/16MM APOLL PS TIB INS SZ3/19MM APOLL PS TIB INS SZ4/9MM APOLLO PS TIB INS SZ4/11MM APOLL PS TIB INS SZ4/13MM APOLL PS TIB INS SZ4/16MM APOLL PS TIB INS SZ4/19MM APOLL PS TIB INS SZ4/22MM APOLL PS TIB INS SZ5/9MM APOLLO PS TIB INS SZ5/19MM APOLL

Quantity: 3638 units

Why Was This Recalled?

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer, Inc.

Zimmer, Inc. has 437 total recalls tracked by RecallDetector.

Related Recalls