Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS Recalled by Zimmer, Inc. Due to Zimmer recently conducted a review of all historic...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer, Inc. directly.
Affected Products
12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX ST VERSYS PRESS FIT LD/FX SZ VERSYS ADVOCATE CEMENTED VERSYS ADVOCATE V-LIGN CE VERSYS CEMENTED FEM STEM VERSYS CT FEM STEM 11X120 VERSYS CT FEM STEM 12X125 VERSYS CT FEM STEM 13X130 VERSYS CT FEM STEM 14X135 VERSYS CT FEM STEM 15X140 VERSYS CT FEM STEM 16X145 VERSYS CT FEM STEM 17X150 VERSYS CEM/REV/CALCAR 13X VERSYS CEM/REV/CALCAR 15X VERSYS CEM/REV/CALCAR 17X VERSYS CEM/REV/CALCAR 19X
Quantity: 7,044,680 total
Why Was This Recalled?
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Zimmer, Inc.
Zimmer, Inc. has 437 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report