Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Vitality Spinal Fixation System Recalled by Zimmer Biomet Spine Inc. Due to Zimmer Biomet is conducting a medical device recall...

Date: May 10, 2017
Company: Zimmer Biomet Spine Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer Biomet Spine Inc. directly.

Affected Products

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

Quantity: 497 units

Why Was This Recalled?

Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer Biomet Spine Inc.

Zimmer Biomet Spine Inc. has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report