Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System Recalled by Zimmer Biomet, Inc. Due to Zimmer Biomet is conducting a voluntary recall for...

Date: April 26, 2017
Company: Zimmer Biomet, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer Biomet, Inc. directly.

Affected Products

Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System, The product is placed into an AlOx coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed product is then placed into a Tyvek/film pouch and that pouch is heat sealed. The fully pouched product is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The labels also include a sterilization indication dot. This system is intended for primary and revision total shoulder arthroplasty, as well as hemi-arthroplasty

Quantity: 16

Why Was This Recalled?

Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.

Where Was This Sold?

This product was distributed to 10 states: AZ, AR, CA, ID, IL, MD, MN, NY, OH, PA

Affected (10 states)Not affected

About Zimmer Biomet, Inc.

Zimmer Biomet, Inc. has 875 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report