Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System Recalled by Zimmer Biomet, Inc. Due to Zimmer Biomet is conducting a voluntary recall for...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer Biomet, Inc. directly.
Affected Products
Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System, The product is placed into an AlOx coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed product is then placed into a Tyvek/film pouch and that pouch is heat sealed. The fully pouched product is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The labels also include a sterilization indication dot. This system is intended for primary and revision total shoulder arthroplasty, as well as hemi-arthroplasty
Quantity: 16
Why Was This Recalled?
Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.
Where Was This Sold?
This product was distributed to 10 states: AZ, AR, CA, ID, IL, MD, MN, NY, OH, PA
About Zimmer Biomet, Inc.
Zimmer Biomet, Inc. has 875 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report