Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ACE Trochanteric Nail System ATN AR SCREW 60MM STERILE ATN Recalled by Zimmer Biomet, Inc. Due to Zimmer Biomet is conducting a medical device recall...

Date: December 20, 2017
Company: Zimmer Biomet, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zimmer Biomet, Inc. directly.

Affected Products

ACE Trochanteric Nail System ATN AR SCREW 60MM STERILE ATN AR SCREW 65MM STERILE ATN AR SCREW 70MM STERILE ATN AR SCREW 75MM STERILE ATN AR SCREW 80MM STERILE ATN AR SCREW 85MM STERILE ATN AR SCREW 90MM STERILE ATN AR SCREW 95MM STERILE ATN AR SCREW 100MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Quantity: 3454

Why Was This Recalled?

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zimmer Biomet, Inc.

Zimmer Biomet, Inc. has 875 total recalls tracked by RecallDetector.