Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ziehm Vision RFD Recalled by Ziehm Imaging Inc Due to Malfunction of the radiation-warning timer and the 3D-specific...

Date: April 15, 2016
Company: Ziehm Imaging Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ziehm Imaging Inc directly.

Affected Products

Ziehm Vision RFD, Mobile C-arm for Mobile Fluoroscopic Imaging

Quantity: 2 devices

Why Was This Recalled?

Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.

Where Was This Sold?

This product was distributed to 17 states: AZ, CA, FL, GA, IL, ME, MA, MI, NV, NJ, NY, PA, RI, TX, UT, VA, WA

Affected (17 states)Not affected

About Ziehm Imaging Inc

Ziehm Imaging Inc has 14 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report