Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope Recalled by Zeiss, Carl Inc Due to Under certain circumstances, the firmware makes it possible...

Date: May 29, 2018
Company: Zeiss, Carl Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zeiss, Carl Inc directly.

Affected Products

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Quantity: 8

Why Was This Recalled?

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Zeiss, Carl Inc

Zeiss, Carl Inc has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report