Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ZAP-X Radiosurgery System Recalled by Zap Surgical Systems Due to If the radiosurgery system triggers a proximity error...

Date: July 25, 2025
Company: Zap Surgical Systems
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Zap Surgical Systems directly.

Affected Products

ZAP-X Radiosurgery System, Model: 300150

Quantity: 17

Why Was This Recalled?

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.