Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Xoran Mini CT Scanner Recalled by Xoran Technologies, Inc. Due to A description of the defect in the product...

Date: July 27, 2010
Company: Xoran Technologies, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Xoran Technologies, Inc. directly.

Affected Products

Xoran Mini CT Scanner

Quantity: 320

Why Was This Recalled?

A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Xoran Technologies, Inc.

Xoran Technologies, Inc. has 6 total recalls tracked by RecallDetector.