Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

stryker Infinity Resection Adjustment Block Recalled by Wright Medical Technology, Inc. Due to Stryker has identified an issue that impacts specific...

Date: March 6, 2024
Company: Wright Medical Technology, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Wright Medical Technology, Inc. directly.

Affected Products

stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty

Quantity: 47 units

Why Was This Recalled?

Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.

Where Was This Sold?

This product was distributed to 2 states: NE, TX

Affected (2 states)Not affected

About Wright Medical Technology, Inc.

Wright Medical Technology, Inc. has 69 total recalls tracked by RecallDetector.

Related Recalls