Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CROSSCHECK(R) Plating system Recalled by Wright Medical Technology, Inc. Due to The minor diameter, major diameter, thread pitch, and...

Date: March 6, 2015
Company: Wright Medical Technology, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Wright Medical Technology, Inc. directly.

Affected Products

CROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Left, REF CCP-HSX1L. Orthopedic use.

Quantity: 51 units

Why Was This Recalled?

The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Wright Medical Technology, Inc.

Wright Medical Technology, Inc. has 69 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report