Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm Recalled by Wright Medical Technology, Inc. Due to These screws have a potential of passing through...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Wright Medical Technology, Inc. directly.
Affected Products
CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014; 3.0 x 16 mm, REF CCP-L3016; 3.0 x 18 mm, REF CCP-L3018; 3.0 x 20 mm, REF CCP-L3020; 3.0 x 22 mm, REF CCP-L3022; 3.0 x 24 mm, REF CCP-L3024; 3.0 x 26 mm, REF CCP-L3026; 3.0 x 28 mm, REF CCP-L3028; 3.0 x 30 mm, REF CCP-L3030; 3.0 x 32 mm, REF CCP-L3032; 3.0 x 34mm, REF CCP-L3034; 3.0 x 36 mm, REF CCP-L3036; 3.0 x 38 mm, REF CCP-L3038; 3.0 x 40 mm, REF CCP-L3040; 3.5 x 8 mm, REF CCP-L3508; 3.5 x 10 mm, REF CCP-L3510; 3.5 x 12 mm, REF CCP-L3512; 3.5 x 14mm, REF CCP-L3514; 3.5 x 16 mm, REF CCP-L3516; 3.5 x 18 mm, REF CCP-L3518; 3.5 x 20 mm, REF CCP-L3520; 3.5 x 22 mm, REF CCP-L3522; 3.5 x 24 mm, REF CCP-L3524; 3.5 x 26 mm, REF CCP-L3526; 3.5 x 28 mm, REF CCP-L3528; 3.5 x 30 mm, REF CCP-L3530; 3.5 x 35 mm, REF CCP-L3532; 3.5 x 34 mm, REF CCP-L3534; 3.5 x 36 mm, REF CCP-L3536; 3.5 x 38 mm, REF CCP-L3538; 3.5 x 40 mm, REF CCP-L3540 The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
Quantity: 107 units
Why Was This Recalled?
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Wright Medical Technology, Inc.
Wright Medical Technology, Inc. has 69 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report