Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with Recalled by Wipro GE Healthcare Private Ltd. Due to GE HealthCare has identified an issue affecting Carescape"...

Date: September 5, 2025
Company: Wipro GE Healthcare Private Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Wipro GE Healthcare Private Ltd. directly.

Affected Products

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

Quantity: 173 units

Why Was This Recalled?

GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Where Was This Sold?

Worldwide distribution.