Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7 Recalled by Wilson-Cook Medical Inc. Due to There are potential risks of the Hemospray powder...

Date: May 22, 2023
Company: Wilson-Cook Medical Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Wilson-Cook Medical Inc. directly.

Affected Products

Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;

Quantity: 63,830 units

Why Was This Recalled?

There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Wilson-Cook Medical Inc.

Wilson-Cook Medical Inc. has 22 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report