HEMO-7 Hemospray Endoscopic Hemostat Recalled by Wilson-Cook Medical Inc. Due to Complaints that the handle and/or activation knob on...

Date: February 4, 2020
Company: Wilson-Cook Medical Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Wilson-Cook Medical Inc. directly.

Affected Products

HEMO-7 Hemospray Endoscopic Hemostat

Quantity: 89,423 devices total

Why Was This Recalled?

Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.