Neo-Vent Infant T-Piece Resuscitation Circuit With Neonatal Mask Recalled by Westmed, Inc. Due to A repair was made to an adapter mold...

Date: January 15, 2019
Company: Westmed, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Westmed, Inc. directly.

Affected Products

Neo-Vent Infant T-Piece Resuscitation Circuit With Neonatal Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0421, Qty:10, Length: 60" (1.5cm), Rx Only.

Quantity: 48 cases

Why Was This Recalled?

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.