Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Propaq LT Series monitors. Model numbers 802LTAN Recalled by Welch Allyn Protocol, Inc Due to Welch Allyn will update Propaq 802 Series Vital...

Date: November 21, 2013
Company: Welch Allyn Protocol, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Welch Allyn Protocol, Inc directly.

Affected Products

The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network.

Quantity: 6003 units total (1715 units in the US and 4288 outside the US)

Why Was This Recalled?

Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.

Where Was This Sold?

This product was distributed nationwide across all 50 states.