Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Replaceable lead set for the Wireless Acquisition Module (WAM) Recalled by WELCH ALLYN, INC/MORTARA Due to Impacted Welch Allyn products may not meet the...

Date: March 18, 2020
Company: WELCH ALLYN, INC/MORTARA
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact WELCH ALLYN, INC/MORTARA directly.

Affected Products

Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following device names: LEAD SET WAM 10 WIRE BANANA AHA GRAY, LEAD SET WAM 10 WIRE BANANA IEC GRAY, RPLCE LD SET WAM/AM12 LIMBS BANA AHA GRY, RPLCE LD SET WAM/AM12 LIMBS BANA IEC GRY, RPLCE LD SET WAM/AM12 V1-V3 BANA AHA GRY, RPLCE LD SET WAM/AM12 C1-C3 BANA IEC GRY, RPLCE LD SET WAM/AM12 V4-V6 BANA AHA GRY, RPLCE LD SET WAM/AM12 C4-C6 BANA IEC GRY, LEAD SET AM15E 13-WIRE BANANA AHA GRAY, LEAD SET AM15E 13-WIRE BANANA IEC GRAY, RPLCE LEAD SET AM15E E2-E4 BANANA AHA GRAY, LEAD SET WAM 10 WIRE CLIPS AHA GRAY, LEAD SET WAM 10 WIRE CLIPS IEC GRAY, RPLCE LD SET WAM/AM12 LIMBS CLIP AHA GRY, RPLCE LD SET WAM/AM12 LIMBS CLIP IEC GRY, RPLCE LD SET WAM/AM12 V1-V3 CLIP AHA GRY, RPLCE LD SET WAM/AM12 C1-C3 CLIP IEC GRY, RPLCE LD SET WAM/AM12 V4-V6 CLIP AHA GRY, RPLCE LD SET WAM/AM12 C4-C6 CLIP IEC GRY, LEAD SET WAM/AM12 10-WIRE BANANA AHA GRY, LEAD SET WAM/AM12 10-WIRE BANANA IEC GRY, LD SET WAM/AM12 10 WRE SHORT AHA CLIP, LD SET WAM/AM12 10 WRE SHORT IEC CLIP.

Quantity: 78,521 units total

Why Was This Recalled?

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About WELCH ALLYN, INC/MORTARA

WELCH ALLYN, INC/MORTARA has 11 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report