Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Welch Allyn KleenSpec Corded Illumination System Recalled by Welch Allyn Inc Due to The firm became aware that the KleenSpec¿ 788...

Date: October 2, 2015
Company: Welch Allyn Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Welch Allyn Inc directly.

Affected Products

Welch Allyn KleenSpec Corded Illumination System, Vaginal Speculum Lighting System, Model Number 78x Series Illuminator, REF 901070, Catalog Numbers 78800, 73410, 73412, 73414, 73416, 78810, 78812, 78814 and 78816, Rx ONLY

Quantity: Domestic: 179,401 units, Foreign: 15,063 units

Why Was This Recalled?

The firm became aware that the KleenSpec¿ 788 Corded Illuminator has the potential to be incorrectly connected to a commonly available AC power cord instead of the illuminator low voltage transformer power cord which is supplied with the illuminator. This may result in a shattered lamp bulb and a potential electric shock hazard.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Welch Allyn Inc

Welch Allyn Inc has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report