Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers Recalled by WASATCH PHOTONICS Due to Two new potential laser safety failure modes have...

Date: August 4, 2025
Company: WASATCH PHOTONICS
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact WASATCH PHOTONICS directly.

Affected Products

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Quantity: 192

Why Was This Recalled?

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Where Was This Sold?

U.S.

About WASATCH PHOTONICS

WASATCH PHOTONICS has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report