Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VS3 Iridium Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS-IR) Product Recalled by Visionsense, Ltd. Due to Broken Optical fiber cable damage can occur through...

Date: October 17, 2022
Company: Visionsense, Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Visionsense, Ltd. directly.

Affected Products

VS3 Iridium Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS-IR) Product Number: 174-0012

Quantity: 205 units: 94 units OUS; 111 units US

Why Was This Recalled?

Broken Optical fiber cable damage can occur through handling or bending and can lead to loss of functionality and heating/melting of the outer cable surface.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Visionsense, Ltd.

Visionsense, Ltd. has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report