Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MRIdian¿ ViewRay¿ Radiation Therapy System Recalled by Viewray Incorporated Due to ViewRay discovered that in the event that an...

Date: August 27, 2015
Company: Viewray Incorporated
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Viewray Incorporated directly.

Affected Products

MRIdian¿ ViewRay¿ Radiation Therapy System, ViewRay¿ Is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Quantity: 4 units

Why Was This Recalled?

ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.

Where Was This Sold?

This product was distributed to 3 states: CA, MO, WI

Affected (3 states)Not affected

About Viewray Incorporated

Viewray Incorporated has 6 total recalls tracked by RecallDetector.

Related Recalls