Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide Recalled by Ventlab LLC Due to Stat-check CO2 indicators are non-functional. The indicator is...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ventlab LLC directly.
Affected Products
STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).
Quantity: 2430
Why Was This Recalled?
Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ventlab LLC
Ventlab LLC has 2 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report