Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BenchMark ULTRA and DISCOVERY ULTRA Instruments Recalled by Ventana Medical Systems Inc Due to Potential for Fluid Leak inside a staining system...

Date: January 20, 2022
Company: Ventana Medical Systems Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ventana Medical Systems Inc directly.

Affected Products

BenchMark ULTRA and DISCOVERY ULTRA Instruments

Quantity: 9214

Why Was This Recalled?

Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.

Where Was This Sold?

This product was distributed nationwide across all 50 states.