Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device Recalled by Varian Medical Systems, Inc. Due to Varian MedicalSystems' X-ray generator Supplier (CPI) has identified...

Date: April 30, 2014
Company: Varian Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Varian Medical Systems, Inc. directly.

Affected Products

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Quantity: 32 units

Why Was This Recalled?

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

Where Was This Sold?

Worldwide Distribution.