Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set. Recalled by Varian Medical Systems Inc Due to Varian has discovered that there is a discrepancy...

Date: May 18, 2015
Company: Varian Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Varian Medical Systems Inc directly.

Affected Products

Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.

Quantity: 46

Why Was This Recalled?

Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat

Where Was This Sold?

Worldwide Distribution.