Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

On-Board Imager - Exact Arm Elbow Motor Shaft (kVD Recalled by Varian Medical Systems, Inc. Due to Varian has identified a failure with the shaft...

Date: March 20, 2014
Company: Varian Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Varian Medical Systems, Inc. directly.

Affected Products

On-Board Imager - Exact Arm Elbow Motor Shaft (kVD, kVS, and MV) with ''Type 01" elbow motors. Product Usage: The On-Board Imager device is an x-ray device used for evaluation of correct patient position in relation to isocenter and verification of treatment fields in relation to anatomical and fiducial landmarks.

Quantity: 2622

Why Was This Recalled?

Varian has identified a failure with the shaft of the Type 01 elbow motor for Exact arms used in the On-Board Imager (OBI) device due to belt over-tensioning.