Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ulrich medical neon3 OCT spinal stabilization Recalled by ulrich medical USA Inc Due to Due to a production error, the weld seam...

Date: December 31, 2019
Company: ulrich medical USA Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ulrich medical USA Inc directly.

Affected Products

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Quantity: 179 units

Why Was This Recalled?

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Where Was This Sold?

This product was distributed to 6 states: AZ, AR, CA, FL, GA, TX

Affected (6 states)Not affected

About ulrich medical USA Inc

ulrich medical USA Inc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report