Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Shanghai United Imaging Healthcare Co. Recalled by UIH Technologies LLC Due to Third-party outer covers can become trapped between patient...

Date: September 12, 2025
Company: UIH Technologies LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact UIH Technologies LLC directly.

Affected Products

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

Quantity: 8 units

Why Was This Recalled?

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Where Was This Sold?

US: OUS:

About UIH Technologies LLC

UIH Technologies LLC has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report