Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Turbett Surgical Container Recalled by Turbett Surgical, Inc. Due to Pod may have weld failures breaching the sterile...

Date: December 5, 2022
Company: Turbett Surgical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Turbett Surgical, Inc. directly.

Affected Products

Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Quantity: 6 units

Why Was This Recalled?

Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Turbett Surgical, Inc.

Turbett Surgical, Inc. has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report