Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Patient transport Shuttle with Trendelenburg quick adjustment function (Jupiter Recalled by Trumpf Medical Systems, Inc. Due to Complaints were received of the table top on...

Date: January 19, 2016
Company: Trumpf Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trumpf Medical Systems, Inc. directly.

Affected Products

Patient transport Shuttle with Trendelenburg quick adjustment function (Jupiter, Mars, Titan, TruSystem and Saturn operating table systems). The shuttle is intended for the following applications: 1. Transfer, transport and storage of exchangeable table tops in the Trumpf Medical TruSystem 7500, JUPITER, and SATURN SM operating table systems 2. Patient transport to an operating table top (without a column) from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system 3. Transporting the TruSystem 7500 SM, JUPITER SM or SATURN SM operating table column with table top (without a patient) within the operating theatre.

Quantity: 1284 affected units

Why Was This Recalled?

Complaints were received of the table top on the transporter (shuttle) tilting down quickly in a maximum Trendelenburg position.

Where Was This Sold?

This product was distributed to 2 states: MA, TX