Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Camera System Recalled by Trumpf Medical Systems, Inc. Due to The radiated emission from the TruVidia" Wireless Receiver...

Date: December 15, 2017
Company: Trumpf Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trumpf Medical Systems, Inc. directly.

Affected Products

Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.

Quantity: 212 devices

Why Was This Recalled?

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.

Where Was This Sold?

This product was distributed to 8 states: GA, IL, IN, PA, SC, TX, UT, VA

Affected (8 states)Not affected

About Trumpf Medical Systems, Inc.

Trumpf Medical Systems, Inc. has 30 total recalls tracked by RecallDetector.

Related Recalls